Troubleshooting

Problem Possibie reason Proposed action
Pipette leaks, Uneven liquid level in tips on multi channel pipette Pipette tip base and tip are not compatibie Change to suitable tip
(e.g. ExpeWM or ExpellPlus™)
Insufficient tip quality Change to suitable tip
(e.g. ExpeWM or ExpellPlus™)
Tip rack bends when
mounting tips on a multi
channel pipette
Change to more solid tip rack
Tip is not firmly attached Press on tip more firmly (avoid hammering pipette into tip
Inaccuracy, imprecision Pipette is not properly calibrated Recalibrate
Sealing ring on the inside of the tip collides with the 0-rings on the tip base Change to tip without internal sealing ring
Top screw on volume controller (marked 51 in the calibration guide) is loose Tighten the screw, recalibrate
O-ringes) on tip base are worn or damaged Change o-ring (s)
Piston seal (inside O-ring) is damaged Dismount tip base (please follow the instructions listed for Lubrication) and replace O-ring
Tip base is loose Tighten the tip base
Liquid remaining in tip after blow-out Tip used is hydrophilic Change to a hydrophobic tip
(e.g. ExpeWM or ExpellPlus™)
Uneven movement of piston Piston seal (inside o-ring) is damaged Dismount tip base (please follow the instructions listed for Lubrication) and replace o-ring
Oil on piston is unevenly
distributed
Set the pipette at maximum volume and operate the plunger fully up and down 3-5 times
Piston needs lubrication Lubricate piston
Deposit on piston Clean the piston and re-Iubricate
 